Designed to enable users to perform quick, easy tests on HART smart control valves. Features built-in test procedures and an intuitive user interface that allows users to quickly and easily perform valve tests, while the valve test quick-check results provide at-a-glance diagnostics helps you make maintenance decisions faster than ever.
Designed to enable users to perform quick, easy tests on HART smart control valves. Features built-in test procedures and an intuitive user interface that allows users to quickly and easily perform valve tests, while the valve test quick-check results provide at-a-glance diagnostics helps you make maintenance decisions faster than ever.
Features
Smart Control Valve Testing is Now Easier Than Ever
The valve testing loop calibrator is designed to enable users to perform quick, easy tests on HART smart control valves. Featuring built-in test procedures and an intuitive user interface, it allows users to quickly and easily perform valve tests, while the valve test quick-check results provide at-a-glance diagnostics help you make maintenance decisions faster than ever. The valve health quick-check results let you know whether your valve is in good, marginal or in bad operating condition so you can quickly decide whether additional maintenance is necessary.
Valve Testing and HART Communication in a Precision Loop Calibrator
With the valve testing loop calibrator's built in HART communication function, users can source a 4-20 mA signal to cause the smart control valve to move, while simultaneously interpreting the valve's HART feedback signal to determine whether the valve is moving to the expected position. In addition to positional information, the measured pressure delivered from the valve's internal I/P (which moves the valve) can be determined through the HART communication protocol. It has built-in test procedures that automatically increase and change the mA signal while monitoring the HART position and pressure feedback from the control valve, giving you a better overall picture of valve health at the simple push of a button.
Pre-Configured Valve Tests, At-A-Glance Answers
Valve test routines built into it include:
ValveTrack™ Software Enables Further Analysis and Trending
Valve tests that are logged and recorded to memory in the loop valve tester are available to upload to included ValveTrack™ analysis software.
ValveTrack™ software enables you to:
Saving Time, Getting Answers
In addition, it offers:
Applications
Functions | mA source, mA simulate, mA read, mA read/loop power, and volts read |
Ranges | mA (0 to 24 mA) and Volts (0 to 30 V dc) |
Resolution | 1uA on mA ranges and 1mV on voltage range |
Accuracy | 0.01% ±2 counts all ranges (@ 73.4°F ±41°F [23°C ±5°C]) |
Operating Temperature Range | 14 to 131°F (-10 to 55°C) |
Humidity Range | 10 to 95% non-condensing |
Stability | 20 ppm of F.S./°C from -10 to 18°C and 28 to 55°C |
Display | 128 x 64 pixels, LCD graphic w/backlight, .34 in high digits |
Power | 6 x AAA alkaline |
Battery Life | ≥40 hours continuous use (measure mode using alkaline) |
Loop Compliance Voltage | 24 V dc @ 20 mA |
Voltage Protection | 240 V ac |
Overload Current Protection | 28 mA dc |
EMC | EN61326 Annex A (portable instruments) |
Dimensions | 6 x 3.6 x 1.3" (15 x 9 x 3 cm) |
Weight | 0.3 kg (9.5 oz) |
Warranty | Three year |
Click on a category to view a selection of compatible accessories with the Fluke 710 mA Loop Valve Tester.
Functions | mA source, mA simulate, mA read, mA read/loop power, and volts read |
Ranges | mA (0 to 24 mA) and Volts (0 to 30 V dc) |
Resolution | 1uA on mA ranges and 1mV on voltage range |
Accuracy | 0.01% ±2 counts all ranges (@ 73.4°F ±41°F [23°C ±5°C]) |
Operating Temperature Range | 14 to 131°F (-10 to 55°C) |
Humidity Range | 10 to 95% non-condensing |
Stability | 20 ppm of F.S./°C from -10 to 18°C and 28 to 55°C |
Display | 128 x 64 pixels, LCD graphic w/backlight, .34 in high digits |
Power | 6 x AAA alkaline |
Battery Life | ≥40 hours continuous use (measure mode using alkaline) |
Loop Compliance Voltage | 24 V dc @ 20 mA |
Voltage Protection | 240 V ac |
Overload Current Protection | 28 mA dc |
EMC | EN61326 Annex A (portable instruments) |
Dimensions | 6 x 3.6 x 1.3" (15 x 9 x 3 cm) |
Weight | 0.3 kg (9.5 oz) |
Warranty | Three year |
Click on a category to view a selection of compatible accessories with the Fluke 710 mA Loop Valve Tester.