This calibrator is designed to simplify the pressure calibration process and provide faster, more accurate test results. Combined with the best in class calibration management software DPCTrack2 creates a complete documenting pressure calibration solution for field applications.
This calibrator is designed to simplify the pressure calibration process and provide faster, more accurate test results. Combined with the best in class calibration management software DPCTrack2 creates a complete documenting pressure calibration solution for field applications.
Combine a Fluke 729 Automatic Pressure Calibrator with Fluke 750 SW DPC/TRACK2 Software for a complete documenting pressure calibration solution.
The Automatic Pressure Calibrator has been designed specifically with process technicians in mind to simplify the pressure calibration process and provide faster, more accurate test results. Technicians know that calibrating pressure can be a time-consuming task, but the 729 makes it easier than ever with an internal electric pump that provides automatic pressure generation and regulation in an in an easy-to-use, rugged, portable package.
Fluke DPC/TRACK2 is a calibration management software database that makes it easy to manage your instrumentation, create and schedule tests, load and unload calibrators, print a variety of standard reports, and manage calibration data. You can also print standard reports automatically. The software assembles pre-formatted reports from your database files, saving time and reducing errors. Reports include calibration certificates, instruments due for calibration, inventory characteristics, calibration histories, calibration procedures, and traceability to instruments touched.
Features
Applications
Pressure Specification | |
1-year Specification | 0.02 % of full scale |
Control Specification | 0.005 % full scale minimum |
Temperature Compensation | 15°C to 35°C (59°F to 95°F) to rated accuracy Note: For temperatures from -10°C to +15 °C and 35°C to 50°C, add 0.04 % of full scale |
Electrical Specification | |
Ranges | 0 mA to 24 mA / 0 V dc to 30 V dc |
Resolution | 1 µA / 1 mV |
Accuracy | 0.01 % ±2 LSD all ranges (at 23°C ±5°C) |
Loop Compliance Voltage | 24 V dc @ 20 mA |
Mechanical Specification | |
Size (H x W x L) | 7.0 cm x 27.9 cm x 17.3 cm (2.75 in x 11.0 in x 6.8 in) |
Weight | 2.95 kg (6.5 lb) |
Click on a category to view a selection of compatible accessories with the Fluke FLUKE-729/750SW BU 729 Automatic Pressure Calibrator and DPC/TRACK2 Software Bundle.
Pressure Specification | |
1-year Specification | 0.02 % of full scale |
Control Specification | 0.005 % full scale minimum |
Temperature Compensation | 15°C to 35°C (59°F to 95°F) to rated accuracy Note: For temperatures from -10°C to +15 °C and 35°C to 50°C, add 0.04 % of full scale |
Electrical Specification | |
Ranges | 0 mA to 24 mA / 0 V dc to 30 V dc |
Resolution | 1 µA / 1 mV |
Accuracy | 0.01 % ±2 LSD all ranges (at 23°C ±5°C) |
Loop Compliance Voltage | 24 V dc @ 20 mA |
Mechanical Specification | |
Size (H x W x L) | 7.0 cm x 27.9 cm x 17.3 cm (2.75 in x 11.0 in x 6.8 in) |
Weight | 2.95 kg (6.5 lb) |