2-, 3- or 4-wire connection programmable temperature transmitter with cable resistance compensation possible in 2-wire system.
2-, 3- or 4-wire connection programmable temperature transmitter with cable resistance compensation possible in 2-wire system.
This programmable temperature transmitter is a two wire transmitter with an analog output. It has measurement input for resistance thermometers (RTD) in 2, 3 or 4 wire connection, thermocouples, resistance and voltage. Setting up of the transmitter is done using the 440-CABLEKIT. These small units can be mounted in Pyromation DIN (Form B) heads or they can be used for surface mounting by using a 35mm DIN rail mounting clip.
Features
Applications
PC programmable temperature head transmitter for converting various input signals into an scalable (4 to 20) mA analog output signal
Connection Type | 2-, 3- or 4-wire connection cable resistance compensation possible in 2-wire system (0 to 20Ω) |
Output Signal | 4 to 20mA or 20 to 4mA |
Measurement Range | |
---|---|
Pt100 RTD | -200 to 850°C (-328 to 1562°F) |
Pt500 RTD | -200 to 250°C (-328 to 482°F) |
Pt1000 RTD | -200 to 250°C (-328 to 482°F) |
Ni100 RTD | -60 to 180°C (-76 to 356°F) |
Ni500 RTD | -60 to 150°C (-76 to 302°F) |
Ni1000 RTD | -60 to 150°C (-76 to 302°F) |
Resistance | 10 to 100Ω, 10 to 2000Ω |
Type B TC | 0 to 1820°C (32 to 3308°F) |
Type C TC | 0 to 2320°C (32 to 4208°F) |
Type D TC | 0 to 2495°C (32 to 4523°F) |
Type E TC | -200 to 915°C (-328 to 1679°F) |
Type J TC | -200 to 1200°C (-328 to 2192°F) |
Type K TC | -200 to 1372°C (-328 to 2501°F) |
Type L TC | -200 to 900°C (-328 to 1652°F) |
Type N TC | -270 to 1300°C (-454 to 2372°F) |
Type R TC | 0 to 1768°C (32 to 3214°F) |
Type S TC | 0 to 1768°C (32 to 3214°F) |
Type T TC | -200 to 400°C (-328 to 752°F) |
Type U TC: | -200 to 600°C (-328 to 1112°F) |
Voltage | -10 to 100mV |
Accuracy | |
Pt100, Ni100 | 0.2°C or 0.08% |
Pt500, Ni500 | 0.5°C or 0.20% |
Pt1000, Ni1000 | 0.3°C or 0.12% |
Resistance | ±0.1Ω or 0.08%: 10 to 400Ω; ±1.5Ω or 0.12%: 10 to 2000Ω |
Type K, J, T, E, L, U | 0.5°C or 0.08% |
Type N, C, D | 1.0°C or 0.08% |
Type S, B, R | 2.0°C or 0.08% |
Voltage | ±20uV or 0.08% |
Click on a category to view a selection of compatible accessories with the Pyromation 441-00 Universal Temperature Transmitter, Unconfigured.
Connection Type | 2-, 3- or 4-wire connection cable resistance compensation possible in 2-wire system (0 to 20Ω) |
Output Signal | 4 to 20mA or 20 to 4mA |
Measurement Range | |
---|---|
Pt100 RTD | -200 to 850°C (-328 to 1562°F) |
Pt500 RTD | -200 to 250°C (-328 to 482°F) |
Pt1000 RTD | -200 to 250°C (-328 to 482°F) |
Ni100 RTD | -60 to 180°C (-76 to 356°F) |
Ni500 RTD | -60 to 150°C (-76 to 302°F) |
Ni1000 RTD | -60 to 150°C (-76 to 302°F) |
Resistance | 10 to 100Ω, 10 to 2000Ω |
Type B TC | 0 to 1820°C (32 to 3308°F) |
Type C TC | 0 to 2320°C (32 to 4208°F) |
Type D TC | 0 to 2495°C (32 to 4523°F) |
Type E TC | -200 to 915°C (-328 to 1679°F) |
Type J TC | -200 to 1200°C (-328 to 2192°F) |
Type K TC | -200 to 1372°C (-328 to 2501°F) |
Type L TC | -200 to 900°C (-328 to 1652°F) |
Type N TC | -270 to 1300°C (-454 to 2372°F) |
Type R TC | 0 to 1768°C (32 to 3214°F) |
Type S TC | 0 to 1768°C (32 to 3214°F) |
Type T TC | -200 to 400°C (-328 to 752°F) |
Type U TC: | -200 to 600°C (-328 to 1112°F) |
Voltage | -10 to 100mV |
Accuracy | |
Pt100, Ni100 | 0.2°C or 0.08% |
Pt500, Ni500 | 0.5°C or 0.20% |
Pt1000, Ni1000 | 0.3°C or 0.12% |
Resistance | ±0.1Ω or 0.08%: 10 to 400Ω; ±1.5Ω or 0.12%: 10 to 2000Ω |
Type K, J, T, E, L, U | 0.5°C or 0.08% |
Type N, C, D | 1.0°C or 0.08% |
Type S, B, R | 2.0°C or 0.08% |
Voltage | ±20uV or 0.08% |
Click on a category to view a selection of compatible accessories with the Pyromation 441-00 Universal Temperature Transmitter, Unconfigured.