This Thermocouple Calibrator is capable of sourcing and measuring 8 different types of thermocouples. The R2810 features a large easy-to-read LCD display, a basic accuracy of 0.05% and includes the essentials to test and calibrate your temperature sensors.
This Thermocouple Calibrator is capable of sourcing and measuring 8 different types of thermocouples. The R2810 features a large easy-to-read LCD display, a basic accuracy of 0.05% and includes the essentials to test and calibrate your temperature sensors.
This thermocouple calibrator is capable of sourcing and measuring 8 different types of thermocouples. The R2810 features an easy-to-read LCD display, basic accuracy of 0.05% and includes the essentials to test and calibrate your temperature sensors.
Features
Measure Function | ||||
Output | Range | Output Range | Resolution | Accuracy |
DCV | 100mV | -10 to 110mV | 0.01mV | 0.05% rdg. +30µV |
TC | R | -40 to 1760°C | 1°C | ±0.05% rdg. +3°C (≤100°C); ±0.05% rdg. +2°C (>100°C) |
S | -20 to 1760°C | 1°C | ||
B | 400 to 1800°C | 1°C | ±0.05% rdg. +3°C (400 to 600°C); ±0.05% rdg. +2°C (>600°C) |
|
E | -200 to 1000°C | 0.1°C | ±0.05% rdg. +2°C (≤-100°C); ±0.05% rdg. +1°C (>-100°C) |
|
K | -200 to 1370°C | 0.1°C | ||
J | -200 to 1200°C | 0.1°C | ||
T | -200 to 400°C | 0.1°C | ||
N | -200 to 1300°C | 0.1°C | ||
Source Function | ||||
Output | Range | Output Range | Resolution | Accuracy |
DCV | 100mV | -10 to 110mV | 0.01mV | 0.05% rdg. +30µV |
1V | -0.1 to 1.1V | 0.1mV | 0.05% rdg. +0.3mV | |
TC | R | -40 to 1760°C | 1°C | ±0.05% rdg. +3°C (≤100°C); ±0.05% rdg. +2°C (>100°C) |
S | -20 to 1760°C | 1°C | ||
B | 400 to 1800°C | 1°C | ±0.05% rdg. +3°C (400 to 600°C); ±0.05% rdg. +2°C (>600°C) |
|
E | -200 to 1000°C | 0.1°C | ±0.05% rdg. +2°C (≤-100°C); ±0.05% rdg. +1°C (>-100°C) |
|
K | -200 to 1370°C | 0.1°C | ||
J | -200 to 1200°C | 0.1°C | ||
T | -200 to 400°C | 0.1°C | ||
N | -200 to 1300°C | 0.1°C |
General Specifications | ||||
Display | 6-Digit LCD | |||
Zero Adjustment Button | Yes | |||
Kick Stand | Yes | |||
Power Supply | 2 AA Batteries | |||
Cold Junction Compensation | Yes | |||
Auto shut off | Yes (after 15 mins/off) | |||
Low Battery Indicator | Yes | |||
Replaceable Test Leads | Yes | |||
Product Certifications | CE | |||
Operating Temperature | 32 to 122°F (0 to 50°C) | |||
Operating Humidity Range | 0 to 85% | |||
Storage Temperature | 14 to 122°F (-10 to 50°C) | |||
Dimensions | 7.1 x 3.5 x 1.9" (180 x 90 x 47mm) | |||
Weight | 8.2oz (500g) |
The REED R2810 is used primarily for calibrating devices that use thermocouples as an input signal such as meters, controllers, transmitters, data loggers and recorders. The accuracy of thermocouple temperatures are an important factor in many applications including: chemical production, petroleum refineries, manufacturing, power production, steel and iron industry. Calibration is performed by measuring/verifying the voltage output of the thermocouple under test.
Click on a category to view a selection of compatible accessories with the REED R2810 Thermocouple Calibrator.
Measure Function | ||||
Output | Range | Output Range | Resolution | Accuracy |
DCV | 100mV | -10 to 110mV | 0.01mV | 0.05% rdg. +30µV |
TC | R | -40 to 1760°C | 1°C | ±0.05% rdg. +3°C (≤100°C); ±0.05% rdg. +2°C (>100°C) |
S | -20 to 1760°C | 1°C | ||
B | 400 to 1800°C | 1°C | ±0.05% rdg. +3°C (400 to 600°C); ±0.05% rdg. +2°C (>600°C) |
|
E | -200 to 1000°C | 0.1°C | ±0.05% rdg. +2°C (≤-100°C); ±0.05% rdg. +1°C (>-100°C) |
|
K | -200 to 1370°C | 0.1°C | ||
J | -200 to 1200°C | 0.1°C | ||
T | -200 to 400°C | 0.1°C | ||
N | -200 to 1300°C | 0.1°C | ||
Source Function | ||||
Output | Range | Output Range | Resolution | Accuracy |
DCV | 100mV | -10 to 110mV | 0.01mV | 0.05% rdg. +30µV |
1V | -0.1 to 1.1V | 0.1mV | 0.05% rdg. +0.3mV | |
TC | R | -40 to 1760°C | 1°C | ±0.05% rdg. +3°C (≤100°C); ±0.05% rdg. +2°C (>100°C) |
S | -20 to 1760°C | 1°C | ||
B | 400 to 1800°C | 1°C | ±0.05% rdg. +3°C (400 to 600°C); ±0.05% rdg. +2°C (>600°C) |
|
E | -200 to 1000°C | 0.1°C | ±0.05% rdg. +2°C (≤-100°C); ±0.05% rdg. +1°C (>-100°C) |
|
K | -200 to 1370°C | 0.1°C | ||
J | -200 to 1200°C | 0.1°C | ||
T | -200 to 400°C | 0.1°C | ||
N | -200 to 1300°C | 0.1°C |
General Specifications | ||||
Display | 6-Digit LCD | |||
Zero Adjustment Button | Yes | |||
Kick Stand | Yes | |||
Power Supply | 2 AA Batteries | |||
Cold Junction Compensation | Yes | |||
Auto shut off | Yes (after 15 mins/off) | |||
Low Battery Indicator | Yes | |||
Replaceable Test Leads | Yes | |||
Product Certifications | CE | |||
Operating Temperature | 32 to 122°F (0 to 50°C) | |||
Operating Humidity Range | 0 to 85% | |||
Storage Temperature | 14 to 122°F (-10 to 50°C) | |||
Dimensions | 7.1 x 3.5 x 1.9" (180 x 90 x 47mm) | |||
Weight | 8.2oz (500g) |
The REED R2810 is used primarily for calibrating devices that use thermocouples as an input signal such as meters, controllers, transmitters, data loggers and recorders. The accuracy of thermocouple temperatures are an important factor in many applications including: chemical production, petroleum refineries, manufacturing, power production, steel and iron industry. Calibration is performed by measuring/verifying the voltage output of the thermocouple under test.
Click on a category to view a selection of compatible accessories with the REED R2810 Thermocouple Calibrator.