Appropriate for any oscilloscope having a high-impedance BNC input, this AC/DC probe has high bandwidth, accurate waveform capture, measurement accuracy of up to 1%, and low circuit insertion loss. Ideal for design and test of switching power supplies, motor drives, electric vehicles, and uninterruptible power supplies.
Appropriate for any oscilloscope having a high-impedance BNC input, this AC/DC probe has high bandwidth, accurate waveform capture, measurement accuracy of up to 1%, and low circuit insertion loss. Ideal for design and test of switching power supplies, motor drives, electric vehicles, and uninterruptible power supplies.
Powered by an external universal power supply and a set of universal plug adaptors.
Features
Applications
Bandwidth (-3 dB) | DC to 12 MHz |
Continuous Maximum Input | 150 A RMS |
Maximum Peak Input (≤10 µs) | 300 A peak |
Ranges | 30 to 150 A |
Overload | 30 to ≥30 A 150 to ≥300 A |
Current Transfer Ratio | 30 A: 1 V/A 150 A: 0.01 V/A |
Oscilloscope Input Scaling | 30 A: x 1 150 A: x 10 |
Minimum Measurable Current | 30 A: 5 mA 150 A: 50 mA |
Amplitude Accuracy (DC, 45 to 66 Hz) | 30 A: ±1%, ±10 mA 150 A: ±1%, ±100 mA |
Maximum Measurement Conductor Diameter | 0.2" (5 mm) |
Power Supply | DC 12 V/1 A (supplied) |
Temperature and Humidity | Operating: 0 to 104°F (0 to 40°C), ≤80% Storage: 14 to 122°F (-10 to 50°C), ≤80% |
Dimensions | Probe head: 6.93 x 1.56 x 0.71" (176 x 39.5 x 18 mm) Probe head to control box cable length: 3.28' (1 m) |
Weight | 0.52 lbs (235 g) |
Bandwidth (-3 dB) | DC to 12 MHz |
Continuous Maximum Input | 150 A RMS |
Maximum Peak Input (≤10 µs) | 300 A peak |
Ranges | 30 to 150 A |
Overload | 30 to ≥30 A 150 to ≥300 A |
Current Transfer Ratio | 30 A: 1 V/A 150 A: 0.01 V/A |
Oscilloscope Input Scaling | 30 A: x 1 150 A: x 10 |
Minimum Measurable Current | 30 A: 5 mA 150 A: 50 mA |
Amplitude Accuracy (DC, 45 to 66 Hz) | 30 A: ±1%, ±10 mA 150 A: ±1%, ±100 mA |
Maximum Measurement Conductor Diameter | 0.2" (5 mm) |
Power Supply | DC 12 V/1 A (supplied) |
Temperature and Humidity | Operating: 0 to 104°F (0 to 40°C), ≤80% Storage: 14 to 122°F (-10 to 50°C), ≤80% |
Dimensions | Probe head: 6.93 x 1.56 x 0.71" (176 x 39.5 x 18 mm) Probe head to control box cable length: 3.28' (1 m) |
Weight | 0.52 lbs (235 g) |