This high performance AC Current Probe provides an accurate voltage output proportional to the measured current (1mV/A, 10mV/A, or 100mV/A).
This high performance AC Current Probe provides an accurate voltage output proportional to the measured current (1mV/A, 10mV/A, or 100mV/A).
This voltage output enables instruments without current ranges to display or log currents on an AC voltage range. The high quality magnetic core and uniform windings provide for accurate measurements up to 30AAC/300AAC/3000AAC. The "squared" jaw permits multiple conductor or bus bar positioning. Provide excellent TRMS capabilities and reliable signal reproduction.
Features
Applications
Nominal Range | 30A, 300A, 3000A |
Measurement Range | 1 to 3200A |
Transformation Ratio | Voltage output |
Output Signal | 100mV AC/A AC (3V at 30A) 90A Peak 10mV AC/A AC (3V at 300A) 900A Peak 1mV AC/A AC (3V at 3000A) 9000A Peak |
Accuracy (30A Range) | 1.5A: 2% of Reading ± 10mV (0.5A) 6A: 2% of Reading ± 10mV (0.5A) 30A: 2% of Reading ± 10mV (0.5A) |
Phase Shift (30A Range) | 1.5A: 15° 6A: 7° 30A: 5° |
Accuracy (300A Range) | 15A: 2% of Reading ± 2mV (0.2mA) 60A: 2% of Reading ± 2mV (0.2mA) 300A: 2% of Reading ± 2mV (0.2mA) |
Phase Shift (300A Range) | 15A: 3° 60A: 1.5° 300A: 1 |
Accuracy (3000A Range) | 150A: 2% of Reading ± 0.5mV 600A: 2% of Reading ± 0.5mV 3000A: 2% of Reading ± 0.5mV |
Phase Shift (3000A Range) | 150A: 1.5° 600A: 1° 3000A: 0.5° |
Overload | 3200A for 5 min |
Frequency Range | 30 to 5kHz Current derating above 200Hz on the 3000A range |
Load Impedance | >100kHz |
Working/Common Mode Voltage | 600V AC |
Output Termination | Jack |
Nominal Range | 30A, 300A, 3000A |
Measurement Range | 1 to 3200A |
Transformation Ratio | Voltage output |
Output Signal | 100mV AC/A AC (3V at 30A) 90A Peak 10mV AC/A AC (3V at 300A) 900A Peak 1mV AC/A AC (3V at 3000A) 9000A Peak |
Accuracy (30A Range) | 1.5A: 2% of Reading ± 10mV (0.5A) 6A: 2% of Reading ± 10mV (0.5A) 30A: 2% of Reading ± 10mV (0.5A) |
Phase Shift (30A Range) | 1.5A: 15° 6A: 7° 30A: 5° |
Accuracy (300A Range) | 15A: 2% of Reading ± 2mV (0.2mA) 60A: 2% of Reading ± 2mV (0.2mA) 300A: 2% of Reading ± 2mV (0.2mA) |
Phase Shift (300A Range) | 15A: 3° 60A: 1.5° 300A: 1 |
Accuracy (3000A Range) | 150A: 2% of Reading ± 0.5mV 600A: 2% of Reading ± 0.5mV 3000A: 2% of Reading ± 0.5mV |
Phase Shift (3000A Range) | 150A: 1.5° 600A: 1° 3000A: 0.5° |
Overload | 3200A for 5 min |
Frequency Range | 30 to 5kHz Current derating above 200Hz on the 3000A range |
Load Impedance | >100kHz |
Working/Common Mode Voltage | 600V AC |
Output Termination | Jack |