The MN352 is a high accuracy and low phase shift AC current probe with voltage output. It is designed for use with DMMs or current measuring instruments with voltage inputs. The voltage output is proportional to the current measured (through the jaws). The MN352 has a jack output that accepts 4mm banana plugs.
Features
Applications
Nominal Range | 150A |
Measurement Range | 0.1 to 150A |
Transformation Ratio | Voltage output |
Output Signal | 10mV/A (1V at 100A AC) |
Accuracy (150A Range) | 0.1 to 1A: ≤3% of Reading ± 200μV 1 to 80A: 1% of Reading ± 0.2mV 80 to 150A: 1.5% of Reading ± 0.2mV |
Phase Shift (150A Range) | 1 to 20A: ≤3.0° 20 to 100A: ≤2.0° 100 to 150A: ≤2.5° |
Overload | 240A for 10 min ON, 30 min OFF |
Frequency Range | 40 to 10kHz |
Limit Operating | 200A permanently to 1kHz Conditions Derating above 3kHz: 200A x (1/0.333F), F in kHz |
Crest Factor | 3 at 150A Peak with an error (due to CF) of 3% |
Working/Common Mode Voltage | 600Vrms |
Output Termination | Jack |
Nominal Range | 150A |
Measurement Range | 0.1 to 150A |
Transformation Ratio | Voltage output |
Output Signal | 10mV/A (1V at 100A AC) |
Accuracy (150A Range) | 0.1 to 1A: ≤3% of Reading ± 200μV 1 to 80A: 1% of Reading ± 0.2mV 80 to 150A: 1.5% of Reading ± 0.2mV |
Phase Shift (150A Range) | 1 to 20A: ≤3.0° 20 to 100A: ≤2.0° 100 to 150A: ≤2.5° |
Overload | 240A for 10 min ON, 30 min OFF |
Frequency Range | 40 to 10kHz |
Limit Operating | 200A permanently to 1kHz Conditions Derating above 3kHz: 200A x (1/0.333F), F in kHz |
Crest Factor | 3 at 150A Peak with an error (due to CF) of 3% |
Working/Common Mode Voltage | 600Vrms |
Output Termination | Jack |