Flexible AC current probe composed of a flexible sensor and an electronic module. The flexible sensor permits measurements on conductors where standard clamp-on probes cannot be used.
Flexible AC current probe composed of a flexible sensor and an electronic module. The flexible sensor permits measurements on conductors where standard clamp-on probes cannot be used.
Composed of an electronic module and flexible sensor, which permits measurements on conductors where standard clamp-on probes cannot be used. The flexibility makes it easy for the sensor to be installed in tight spaces, around cable bundles, around wide or large bus bars, or even wrapped around irregular shapes. The Shape Memory feature enables the user to pre-shape the sensor before inserting it between or around conductors. This feature facilitates positioning the sensor around the conductor, enhances user safety, and alleviates the drooping effect associated with other flexible sensors.
Features
Applications
Electrical Specifications | |
Measurement Range | 5 to 3000A |
Output Signal | 10mV/A & 1mV/A |
Accuracy | 1% of Reading |
Crest Factor | 3 |
Phase Shift @ 50/60Hz | 0.7° |
Frequency Range | 10 to 20,000Hz (intentionally limited for optimum performance) |
Residual Noise | 500mA |
Working Voltage | 1000Vrms max. |
Influence of Frequency | Add 1% to the reference |
Influence of Adjacent Conductor | <2% |
Influence of Conductor Position in Jaw | <4% |
Mechanical Specifications | |
Output | Two 4mm safety banana jacks, standard 3/4" (19mm) spacing |
Battery | 9V Alkaline |
Low Battery | Green LED Off when battery reaches 7V |
Overload Indication | Red LED indicates input greater than selected range |
Battery Life | 150 hours typical (continuous use) |
Sensor Dimensions | 36" |
Electronic Module Dimensions | 4.9 x 2.5 x 1.1" (124 x 64 x 28mm) |
Cable Length | 6.5ft (2m) |
Drop Test | 1.0m on 38mm of oak or concrete Test according to IEC1010 |
Vibration | Test per IEC 68-2-6 Frequency Range: 10Hz to 55Hz Amplitude: 0.15mm |
Mechanical Shock | 100G, Test per IEC 68-2-32 |
Environmental Specification | |
Case Protection | Sensor: IP65 per IEC529; Electronic Module: IP40 |
Operating Temperature | 14 to 131°F (-10 to 55°C) |
Storage Temperature | -40 to 158°F (40 to 70°C) |
Influence of Temperature | Sensor: -10°C to 100°C: <0.5% per 10°C Module: -10°C to 55°C: <0.5% per 10°C |
Operative Relative Humidity | 10 to 30°C 85 ±5% RH without condensation 40 to 50°C 45 ±5% RH (without condensation) |
Altitude | Operating: 0 to 2000m; non-operating: 0 to 12000m |
Safety Reference Specifications | |
Safety Ratings | EN 61010, 1000V, Cat. III, Pollution Degree 2 |
Agency Approvals | CE Mark |
Electrical Specifications | |
Measurement Range | 5 to 3000A |
Output Signal | 10mV/A & 1mV/A |
Accuracy | 1% of Reading |
Crest Factor | 3 |
Phase Shift @ 50/60Hz | 0.7° |
Frequency Range | 10 to 20,000Hz (intentionally limited for optimum performance) |
Residual Noise | 500mA |
Working Voltage | 1000Vrms max. |
Influence of Frequency | Add 1% to the reference |
Influence of Adjacent Conductor | <2% |
Influence of Conductor Position in Jaw | <4% |
Mechanical Specifications | |
Output | Two 4mm safety banana jacks, standard 3/4" (19mm) spacing |
Battery | 9V Alkaline |
Low Battery | Green LED Off when battery reaches 7V |
Overload Indication | Red LED indicates input greater than selected range |
Battery Life | 150 hours typical (continuous use) |
Sensor Dimensions | 36" |
Electronic Module Dimensions | 4.9 x 2.5 x 1.1" (124 x 64 x 28mm) |
Cable Length | 6.5ft (2m) |
Drop Test | 1.0m on 38mm of oak or concrete Test according to IEC1010 |
Vibration | Test per IEC 68-2-6 Frequency Range: 10Hz to 55Hz Amplitude: 0.15mm |
Mechanical Shock | 100G, Test per IEC 68-2-32 |
Environmental Specification | |
Case Protection | Sensor: IP65 per IEC529; Electronic Module: IP40 |
Operating Temperature | 14 to 131°F (-10 to 55°C) |
Storage Temperature | -40 to 158°F (40 to 70°C) |
Influence of Temperature | Sensor: -10°C to 100°C: <0.5% per 10°C Module: -10°C to 55°C: <0.5% per 10°C |
Operative Relative Humidity | 10 to 30°C 85 ±5% RH without condensation 40 to 50°C 45 ±5% RH (without condensation) |
Altitude | Operating: 0 to 2000m; non-operating: 0 to 12000m |
Safety Reference Specifications | |
Safety Ratings | EN 61010, 1000V, Cat. III, Pollution Degree 2 |
Agency Approvals | CE Mark |