This Current Probe is designed for use in utility and industrial environments. The unique ergonomic design allows it to easily clamp onto cables or small bus bars. It is assembled to the highest safety and performance standards including the CE Mark and is UL approved for Canada and United States.
This Current Probe is designed for use in utility and industrial environments. The unique ergonomic design allows it to easily clamp onto cables or small bus bars. It is assembled to the highest safety and performance standards including the CE Mark and is UL approved for Canada and United States.
This range selectable probe offers four switch selectable ranges: 1, 10, 100 or 1000A. This probe has a voltage output proportional to the current measured in the jaw of the probe. The SR759 has an integral 5 ft (1.5m) lead terminated by shrouded banana plugs compatible with DMMs and other instruments. The high accuracy, high sensitivity and low phase shift makes this probe an excellent choice for use in a wide range of power quality applications, including very low or leakage current measurements.
Features
Applications
Nominal Range | 1A, 10A, 100A, 1000A |
Measurement Range | 1mA to 1200A |
Transformation Ratio | Voltage output |
Output Signal | 1A: 1000mV 10A: 100mV 100A: 10mV 1000A: 1mV/A AC |
Accuracy (1A Range) | 0.001 to 0.1A: 3% of Reading ± 1mV 0.1 to 1A: 0.7% of Reading ± 1mV |
Phase Shift (1A Range) | 0.1 to 1A: <10° |
Accuracy (10A Range) | 0.01 to 0.1A: 1% of Reading ± 0.2mV 0.1 to 1A: 0.5% of Reading ± 0.2mV 1 to 10A: <0.5% of Reading |
Phase Shift (10A Range) | 0.1 to 1A: <5° 1 to 10A: <2° |
Accuracy (100A Range) | 0.1 to 1A: 1% of Reading ± 0.2mV 1 to 10A: 0.5% of Reading ± 0.2mV 10 to 100A: 0.3% of Reading |
Phase Shift (100A Range) | 1 to 10A: <2° 10 to 100A: <1° |
Accuracy (1000A Range) | 1 to 10A: 1% of Reading ± 0.2mV 10 to 100A: 0.5% of Reading ± 0.2mV 100 to 1000A: 0.2% of Reading 100 to 1200A: 0.2% of Reading |
Phase Shift (1000A Range) | 1 to 10A: <2° 10 to 100A: <2° 1000 to 1200A: <1° |
Overload | 1.2A Arms continuous 12A Arms continuous 120A Arms continuous 1200A Arms for five min |
Frequency Range | 30 to 3000Hz Current derating above 1kHz using the formula: 1000A x 1F (in kHz) |
Load Impedance | 100kΩ min |
Working/Common Mode Voltage | 600Vrms Cat. III |
Output Termination | 5ft Lead |
Nominal Range | 1A, 10A, 100A, 1000A |
Measurement Range | 1mA to 1200A |
Transformation Ratio | Voltage output |
Output Signal | 1A: 1000mV 10A: 100mV 100A: 10mV 1000A: 1mV/A AC |
Accuracy (1A Range) | 0.001 to 0.1A: 3% of Reading ± 1mV 0.1 to 1A: 0.7% of Reading ± 1mV |
Phase Shift (1A Range) | 0.1 to 1A: <10° |
Accuracy (10A Range) | 0.01 to 0.1A: 1% of Reading ± 0.2mV 0.1 to 1A: 0.5% of Reading ± 0.2mV 1 to 10A: <0.5% of Reading |
Phase Shift (10A Range) | 0.1 to 1A: <5° 1 to 10A: <2° |
Accuracy (100A Range) | 0.1 to 1A: 1% of Reading ± 0.2mV 1 to 10A: 0.5% of Reading ± 0.2mV 10 to 100A: 0.3% of Reading |
Phase Shift (100A Range) | 1 to 10A: <2° 10 to 100A: <1° |
Accuracy (1000A Range) | 1 to 10A: 1% of Reading ± 0.2mV 10 to 100A: 0.5% of Reading ± 0.2mV 100 to 1000A: 0.2% of Reading 100 to 1200A: 0.2% of Reading |
Phase Shift (1000A Range) | 1 to 10A: <2° 10 to 100A: <2° 1000 to 1200A: <1° |
Overload | 1.2A Arms continuous 12A Arms continuous 120A Arms continuous 1200A Arms for five min |
Frequency Range | 30 to 3000Hz Current derating above 1kHz using the formula: 1000A x 1F (in kHz) |
Load Impedance | 100kΩ min |
Working/Common Mode Voltage | 600Vrms Cat. III |
Output Termination | 5ft Lead |